Wheatstone Bridge

“Derive a bridge’s null condition from equal potentials, distinguish balance from an open detector and solve unbalanced bridge networks.”

Learning Objectives

• Derive the balance relation with explicitly labelled bridge arms. • Use a null measurement to determine an unknown resistance. • Solve an unbalanced bridge with finite detector resistance. • Explain the meter-bridge ratio and practical uncertainty sources.

A detector compares two potentials

A Wheatstone bridge has two voltage-divider paths between supply nodes A and C. Call the top path A–B–C, with resistances P on AB and Q on BC. Call the bottom path A–D–C, with R on AD and S on DC. A galvanometer connects B and D. It responds to current in that detector branch. At balance its current is zero because B and D have equal potential. The main arms can still carry current and dissipate power.

PQRSGsupply+−ABCD
Bridge labels used throughout this lesson— The supply is across A and C, positive toward A; the detector G is across B and D. P, Q, R and S refer to connected arms, not arbitrary positions in another drawing.

Derive the null condition before memorising a ratio

Assume a nonzero supply across A and C, positive finite arm resistances and an ideal zero-current null. With I_G = 0, the upper current I_t is the same through P and Q, and the lower current I_b is the same through R and S. Equal potentials at B and D imply equal drops from A: I_tP = I_bR. They also imply equal drops to C: I_tQ = I_bS. Divide the two equations; the currents cancel and give P/Q = R/S. The ratio depends on which physical arms carry the chosen names.

Wheatstone balance and voltage-divider interpretationLaTeX
The voltage-divider formulas apply at null or with the detector disconnected. With a conducting unbalanced detector it loads both dividers, so these simple divider expressions are not generally the final node voltages.

Alternatively, equate P/(P+Q) and R/(R+S). Cross multiplication gives P(R+S) = R(P+Q); cancelling PR leaves PS = RQ. This establishes the same result without choosing branch-current magnitudes. The source’s internal resistance can reduce the actual A–C voltage, but does not change the ideal null ratio because that common voltage cancels. A lower source voltage can reduce sensitivity to a small imbalance, so practical precision is a separate issue.

Measure an unknown arm

Problem
The bridge balances with P = 100 Ω, Q = 50 Ω and R = 60 Ω. Find S and the arm currents for V_AC = 6.0 V.

  1. 1.From P/Q = R/S, solve PS = QR, so S = QR/P = (50 × 60)/100 = 30 Ω.
  2. 2.At balance the upper path is 150 Ω and carries 6/150 = 0.040 A. The lower path is 90 Ω and carries 6/90 = 0.0667 A.
  3. 3.From V_C = 0, V_B = I_tQ = 2.0 V and V_D = I_bS = 2.0 V. The detector has zero current even though total source current is 0.1067 A.

Null is a physical condition, not merely a disconnected meter

An open detector branch carries no current for any B–D voltage, so removing the galvanometer does not prove the bridge is balanced. Conversely, a finite-resistance detector connected at equal potentials carries zero current and does not alter the ideal balance. A real zero reading can also result from a faulty detector or absent supply, so a functioning measurement must test sensitivity near the apparent null. Swapping the supply and detector diagonals preserves the ideal balance condition, but practical sensitivity and instrument constraints can change.

An unbalanced detector changes the circuit

If P/Q differs from R/S, current generally passes through G. Then the two resistors of a divider are no longer in simple series because current can branch at B or D. Keep the detector in the network and use Kirchhoff’s rules. A convenient method assigns node voltages V_B and V_D relative to C = 0 while the ideal source fixes V_A = V_AC. Every resistor current is its signed node-potential difference divided by resistance.

Node equations for a loaded bridgeLaTeX
R_G is the detector resistance. These are outgoing-current sums at B and D. The signed detector current from B to D is (V_B − V_D)/R_G.
A bridge that is not balanced

Problem
Set P = 100 Ω, Q = 10 Ω, R = 60 Ω, S = 5 Ω and R_G = 15 Ω, with V_A = 10 V and V_C = 0. Find the detector current.

  1. 1.The ratios are 100/10 = 10 and 60/5 = 12, so do not impose zero detector current. At B: (V_B − 10)/100 + V_B/10 + (V_B − V_D)/15 = 0. Multiply by 300 to get 53V_B − 20V_D = 30.
  2. 2.At D: (V_D − 10)/60 + V_D/5 + (V_D − V_B)/15 = 0. Multiply by 60 to get −4V_B + 17V_D = 10.
  3. 3.From the second equation V_D = (10 + 4V_B)/17. Substitution in the first gives 901V_B − 200 − 80V_B = 510, so V_B = 710/821 V and V_D = 650/821 V.
  4. 4.I_G = (60/821)/15 = 4/821 A = 4.87 mA from B to D. Upper incoming current is (10 − V_B)/100 ≈ 91.35 mA and outgoing toward C is V_B/10 ≈ 86.48 mA; their difference is the detector current.

A meter bridge makes one ratio geometric

A uniform resistance wire of total length L can provide two adjacent arms. With a sliding contact at length ℓ, their resistances are proportional to ℓ and L − ℓ because they have the same material and area. If those wire arms are P and Q and the other arms are known R and unknown X, the balance relation is ℓ/(L − ℓ) = R/X. Solving gives X = R(L − ℓ)/ℓ. The numerator order changes if the known and unknown arms are swapped; label the connections instead of memorising an orientation-free length formula.

P and Q are two parts of one uniform wireABCP ∝ ℓQ ∝ L − ℓdetector contactOther bridge arms: R and X; null gives P/Q = R/X
Uniform-wire arms in a meter-bridge ratio— This isolates the uniform-wire pair from the full bridge drawn above. The detector’s other end connects to the junction between R and X; the source is across A and C.
Use and question a length ratio

Problem
A 1.00 m uniform wire balances at ℓ = 0.40 m with R = 20 Ω and unknown X in the stated orientation. Find X.

  1. 1.X = R(L − ℓ)/ℓ = 20 × 0.60/0.40 = 30 Ω. Both lengths must use the same units.
  2. 2.A contact near an endpoint makes a small absolute length error a larger fraction of the short arm. Keeping balance away from the ends helps practical sensitivity.
  3. 3.Contact resistance, end connections, wire nonuniformity, self-heating and finite detector sensitivity limit the ideal ratio. Interchanging known and unknown arms is a useful consistency check, not a guarantee of perfect cancellation of errors.

A bridge with a nonideal supply

For an unbalanced bridge, an external series resistor or source internal resistance changes V_AC as well as the total current. First solve the bridge currents in terms of its actual terminal voltage or find its equivalent resistance; then solve the supply circuit. Substituting the source emf directly for V_AC would ignore its series drop. The chapter-wide practice includes such a network. This connects null measurement, ordinary networks and real-source modelling without changing the bridge topology.

Quick Quiz

Quiz

Quick check

With P = AB, Q = BC, R = AD and S = DC, balance requires

Explanation: Equal drops from A to B and A to D, and from B/D to C, give PS = QR.

Quick check

A functioning bridge detector reads null because

Explanation: Equal detector-end potentials give zero current even for finite detector resistance.

Quick check

Removing the detector and seeing no detector current proves

Explanation: An open branch carries no current even when its endpoints have different potentials.

Quick check

At balance, a source internal resistance ideally changes

Explanation: The common supply voltage cancels from the balance relation.

Quick check

An unbalanced finite-resistance detector means the two upper arms are

Explanation: At their common node current can branch through the detector.

Practice Problems

Practice Problems
  1. A bridge balances with P = 200 Ω, Q = 80 Ω and R = 150 Ω. Find S.
  2. At balance with P = Q = 100 Ω and R = S = 200 Ω, V_AC = 12 V. Find total supply current and detector current.
  3. For P = 100 Ω, Q = 100 Ω, R = 100 Ω and S = 300 Ω at V_AC = 8 V, find V_B − V_D with the detector disconnected and predict current direction when it is connected.
  4. A 100 cm meter-bridge wire balances at ℓ = 60 cm with R = 30 Ω and unknown X, using P/Q = R/X. Find X and predict the new ℓ if R and X swap arms.
  5. Explain why a low detector current alone is insufficient evidence of an accurate unknown resistance.

Use P/Q = R/S, so S = QR/P = (80 × 150)/200 = 60 Ω. Check that both ratios equal 2.5. The relation is tied to the A–B–C and A–D–C arm labels in this lesson.

Key Takeaways

Key Takeaways

• A bridge compares potentials at the junctions of two divider paths. • For the stated labels balance requires P/Q = R/S. • Zero detector current at balance does not mean zero arm currents. • An open detector is not proof of equal endpoint potentials. • A connected unbalanced detector loads both dividers. • Kirchhoff or node equations handle finite detector resistance. • A meter bridge uses a uniform wire’s resistance-to-length ratio. • Supply loading and measurement uncertainty must be treated separately from the ideal null condition.

Finished this lesson? Mark it complete to track your progress.