Moving-Coil Galvanometers and Meters

“Connect magnetic torque with deflection, sensitivity, ammeter shunts, voltmeter series resistance and measurement loading.”

Learning Objectives

• Explain the radial-field moving-coil galvanometer and derive its deflection law. • Compare current and voltage sensitivities. • Derive shunt and series-resistance requirements for chosen measurement ranges. • Quantify how meter resistance changes the circuit being measured.

Turn a current into a deflection

A moving-coil galvanometer contains a light coil with N turns and area A per turn, suspended or pivoted so it can rotate. Curved magnetic pole pieces and a cylindrical soft-iron core create a strong approximately radial magnetic field over the operating region. The coil sides experience forces producing a turning couple. A spring provides the opposing torsional torque, and a pointer indicates the settled deflection φ.

A radial field is not one globally uniform vector: its direction changes around the gap. The design keeps the local field in the coil plane for each operating angle, making the effective torque NIAB independent of pointer deflection. This is different from a freely rotating coil in one spatially uniform fixed-direction field, for which torque has a sin θ factor that varies during rotation. The core helps shape and strengthen the field; it is not a replacement for the torque law.

Balance magnetic and restoring torquesLaTeX
k is the torsional constant in N m rad⁻¹; φ is angular deflection in radians; B is the effective radial gap-field magnitude. The law describes static equilibrium within the calibrated operating range.

At zero current the spring returns the pointer toward its neutral position. Reversing current reverses magnetic torque and deflection. A centre-zero detector is useful for seeing either current direction or a bridge balance. During motion, inertia and damping matter; the static balance equation does not describe every instant of the pointer’s transient movement.

Sensitivity is deflection per input

Current and voltage sensitivitiesLaTeX
R_G is the coil resistance for the bare instrument. S_I is in rad A⁻¹ and S_V in rad V⁻¹. For an instrument with added series resistance, use the total input resistance in the voltage-sensitivity expression.

Increasing turns, area or field, or reducing spring stiffness, can increase current sensitivity if the other quantities stay fixed. More turns often mean more wire and a larger coil resistance. Doubling N while also doubling R_G doubles current sensitivity but leaves voltage sensitivity unchanged if A, B and k remain fixed. Sensitivity, measurement range and accuracy are distinct: a sensitive detector can still disturb a circuit or have calibration errors.

Compare two instruments

Problem
Meter 1 has N = 30, A = 3.6 × 10⁻³ m², B = 0.25 T and R_G = 10 Ω. Meter 2 has N = 42, A = 1.8 × 10⁻³ m², B = 0.50 T and R_G = 14 Ω. Their springs have equal k. Compare sensitivities.

  1. 1.S_I2/S_I1 = (42 × 1.8 × 0.50)/(30 × 3.6 × 0.25) = 1.40.
  2. 2.S_V2/S_V1 = (S_I2/S_I1)(R_G1/R_G2) = 1.40 × 10/14 = 1.00.
  3. 3.The second instrument gives more deflection per ampere but the same deflection per applied volt. Its increased resistance offsets its increased current sensitivity.

Ammeter conversion: divert most of the current

Let the galvanometer reach full-scale deflection at current I_g and have resistance R_G. To measure a larger total current I_max, connect a small shunt R_s in parallel with the coil. At full scale, I_g passes through the coil and I_max − I_g through the shunt. Parallel branches have equal voltage. This gives the shunt resistance directly.

Ammeter range and resistanceLaTeX
I_max must exceed I_g. R_A is the complete ammeter’s resistance. It is approximately R_s when R_s is much smaller than R_G.

The complete ammeter goes in series with the branch whose current is being measured. Its low resistance limits the change it makes to that branch. A shunt does not make the coil capable of carrying the full branch current; it routes most current around the coil and the scale is calibrated for the total. An ammeter connected directly across a voltage source can carry a very large unintended current because its input resistance is small.

Make a 1 A meter

Problem
A galvanometer has I_g = 1.0 mA and R_G = 100 Ω. Find the shunt for a 1.0 A full-scale range.

  1. 1.Full-scale coil voltage is I_gR_G = 0.0010 × 100 = 0.10 V. The shunt must carry 0.999 A at the same voltage.
  2. 2.R_s = 0.10/0.999 = 0.1001 Ω. The combined resistance is R_A = 0.1000 Ω.
  3. 3.At full scale the shunt dissipates approximately 0.10 W, so an actual design must use a resistor with suitable tolerance and power rating. The current division is the reason the coil is protected within its intended range.

Voltmeter conversion: draw little current

To measure voltage, place a large resistor R_series in series with the galvanometer and connect the complete combination across the target component. Full scale occurs when the applied voltage V_max drives exactly I_g through both series elements. Their voltage drops add; unlike the ammeter conversion, there is no parallel current division inside this instrument.

Voltmeter range and input resistanceLaTeX
The chosen V_max must be at least the coil full-scale voltage I_gR_G for a nonnegative added resistance. R_V is the complete voltmeter’s input resistance.
Make a 10 V meter

Problem
Use the same 1.0 mA, 100 Ω galvanometer to obtain a 10 V full-scale range.

  1. 1.The required input resistance is R_V = V_max/I_g = 10/0.0010 = 10 000 Ω.
  2. 2.R_series = 10 000 − 100 = 9900 Ω. At 10 V the coil carries only its 1.0 mA full-scale current.
  3. 3.The voltmeter draws current from the measured circuit. A 10 kΩ input may be acceptable in a low-resistance circuit but can produce substantial loading in a high-resistance one.
Ammeter: shunt in parallelGR_sComplete meter in seriesVoltmeter: series resistorGR_seriesComplete meter across targetG includes its coil resistance R_G
Meter conversion connections— Connection topology is schematic. The galvanometer and shunt share both nodes; the voltmeter resistor and galvanometer share one series path.

A meter changes the circuit

The ammeter loading comparison

Problem
An ideal 3.0 V source supplies a 3.0 Ω resistor. Compare a 60 Ω bare galvanometer, that coil shunted by 0.020 Ω, and an ideal ammeter inserted in series.

  1. 1.Without added resistance the current is 3.0/3.0 = 1.00 A. A bare 60 Ω coil changes the total resistance to 63 Ω, so the current becomes 0.0476 A. This is a large loading effect, and may exceed that coil’s safe current.
  2. 2.With the shunt, R_A = (60 × 0.020)/(60.020) = 0.019993 Ω. The current is 3.0/(3.0 + 0.019993) = 0.9934 A.
  3. 3.An ideal zero-resistance ammeter gives 1.00 A. The low-resistance shunt makes the reading much closer to the undisturbed value, but does not remove loading exactly.

A voltmeter likewise changes the voltage it measures. Treat the rest of a linear circuit at the measurement terminals as a source V_th with series resistance R_th. A voltmeter of input R_V reads V_th R_V/(R_th + R_V). If R_V is much larger than R_th, the error is small. For error below 1% relative to V_th, R_V must exceed 99R_th in this simple model. “High resistance” is always relative to the measured circuit.

A galvanometer converts electrical input to a calibrated mechanical indication. Its torque law, spring balance and circuit connections work together. Use the mechanical model to understand deflection, then the circuit model to choose range and quantify loading; neither alone is sufficient.

Quick Quiz

Quiz

Quick check

Why is a radial field useful in a galvanometer?

Explanation: The geometry maintains the effective perpendicular condition for the active coil sides.

Quick check

An ammeter shunt is connected

Explanation: Parallel branches share voltage and divert most current through the low resistance.

Quick check

A voltmeter conversion requires

Explanation: A large series resistance limits the current drawn for a given voltage.

Quick check

Doubling N and R_G at fixed A, B and k makes bare voltage sensitivity

Explanation: S_V is proportional to N/R_G, which stays fixed.

Quick check

To reduce ammeter loading, its total input resistance should be

Explanation: Series resistance changes the branch current; a small input resistance limits that change.

Practice Problems

Practice Problems
  1. For N = 100, A = 2.0 × 10⁻⁴ m², B = 0.20 T and k = 2.0 × 10⁻⁶ N m rad⁻¹, find current sensitivity and deflection at 0.10 mA.
  2. A 50 Ω coil reaches full scale at 2.0 mA. Find a shunt for a 0.50 A range.
  3. Convert that 50 Ω, 2.0 mA coil to a 5.0 V voltmeter.
  4. A 10 V source with series resistance 2.0 kΩ is measured by a 10 kΩ voltmeter. Find the reading and fractional error.
  5. Explain why a more current-sensitive galvanometer need not be more voltage-sensitive.

S_I = NAB/k = (100)(2.0 × 10⁻⁴)(0.20)/(2.0 × 10⁻⁶) = 2000 rad A⁻¹. With I = 1.0 × 10⁻⁴ A, φ = 0.20 rad, about 11.5°. This uses the static linear range.

Key Takeaways

Key Takeaways

• A radial-field galvanometer balances magnetic and spring torques. • Static deflection is φ = (NAB/k)I. • Current and voltage sensitivities are different. • An ammeter uses a parallel shunt and goes in series with the target branch. • A voltmeter uses a series resistor and goes across the target component. • Low ammeter resistance and high voltmeter resistance reduce loading. • A real meter’s range and resistance must be considered together. • The radial field is a designed geometry, not a globally uniform vector.

Finished this lesson? Mark it complete to track your progress.